Case 01 · BPL patient monitor
Stuck on the boot screen. Back to showing vitals.
- It powered on, showed the brand logo, and went no further.
- The team suspected invisible micro-cracks in the solder joints, the kind pump vibration and cuff stress can cause over time.
- They applied flux, reflowed it with a hot-air rework station, and rechecked the capacitors with a multimeter.
- Switched back on, it booted past the logo to live vitals.